Abstract
The aim of this study was to investigate the feasibility of head and neck computed tomography angiography (CTA) using the 80-kV tube voltage and the adaptive iterative dose reduction (AIDR) 3D algorithm in patients with different body mass indexes (BMIs).From November 2016 to January 2017, 128 consecutive patients scheduled for head and neck CTA examinations were randomized into the 100-kV group (n = 60) and the 80-kV group (n = 68). Both groups used the automatic tube current modulation technique and the AIDR 3D algorithm. The patients were further grouped as slender (BMI < 22 kg/m), normal weight (22 kg/m≤BMI < 25 kg/m), and overweight (BMI ≥25 kg/m). The image quality and the radiation dose of each subgroup were analyzed.The images of the head and neck vessels and the brain tissue obtained with 100 kV were all of diagnostic quality. Slender and normal weight patients imaged with 80 kV also produced images of diagnostic quality. However, 80 kV in the overweight patients failed to produce images of diagnostic quality. The radiation dose in the patients imaged with 80 kV was significantly decreased in comparison with those imaged with 100 kV. The effective dose was 0.36 ± 0.06 and 0.41 ± 0.05 mSv in the slender and normal weight patients imaged with 80 kV.Head and neck CTA scanning with 80 kV, automatic tube current modulation, and AIDR 3D algorithm can produce high quality images with reduced radiation dose in slender or normal weight patients.