Characterization and suppression of stripe artifact in velocity-selective magnetization-prepared unenhanced MR angiography

速度选择性磁化准备非增强磁共振血管造影中条纹伪影的特征分析与抑制

阅读:1

Abstract

PURPOSE: To characterize and suppress stripe artifact associated with velocity-selective (VS) magnetization for unenhanced MRA. METHODS: Extended phase graph formalism was used to show that the stripe artifact contains multiples of the fundamental frequency that is determined by the area of unipolar VS gradient. Four VS preparation pulses whose excitation profiles are spatially shifted by quarter the fundamental period of the stripes, were applied alternately. For further suppression of the artifact, k-space data at k(z)  = 0 were averaged over the 4 VS preparations. The proposed schemes were tested in a chicken breast phantom and healthy human subjects. RESULTS: When the standard VS preparation scheme was used, stripe artifact was shown in all the reconstructed images and appeared as artifactual peaks in k-space that corresponded to the first and second order harmonics of the fundamental frequency. Alternate application of the 4 phase-shifted VS preparation pulses suppressed the stripes, but not completely, as evidenced by residual erroneous peaks in k-space. After the k-space averaging, the stripe artifact was nearly eliminated. CONCLUSION: Stripe artifact in VS-MRA consists of multiples of the fundamental frequency and can be effectively suppressed through alternate application of phase-shifted VS preparations along with k-space averaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。