B1 Field inhomogeneity correction for qDESS T2 mapping: application to rapid bilateral knee imaging

B1 qDESS T2 mapping 的场不均匀性校正:在快速双侧膝关节成像中的应用

阅读:1

Abstract

PURPOSE: T2 mapping is a powerful tool for studying osteoarthritis (OA) changes and bilateral imaging may be useful in investigating the role of between-knee asymmetry in OA onset and progression. The quantitative double-echo in steady-state (qDESS) can provide fast simultaneous bilateral knee T2 and high-resolution morphometry for cartilage and meniscus. The qDESS uses an analytical signal model to compute T2 relaxometry maps, which require knowledge of the flip angle (FA). In the presence of B1 inhomogeneities, inconsistencies between the nominal and actual FA can affect the accuracy of T2 measurements. We propose a pixel-wise B1 correction method for qDESS ρc = 0.69). Without the B1 correction, T2 values showed a linear dependence with B1. The linear coefficient decreased using the B1 correction (from 24.3 ± 1.6 ms to 4.1 ± 1.8) and the correlation was not statistically significant after the application of the Bonferroni correction (p value > 0.01). CONCLUSION: The study showed that B1 correction could mitigate variations driven by the sensitivity of the qDESS T2 mapping method to B1, therefore, increasing the sensitivity to detect real biological changes. The proposed method may improve the robustness of bilateral qDESS T2 mapping, allowing for an accurate and more efficient evaluation of OA pathways and pathophysiology through longitudinal and cross-sectional studies.