The Assessment of Retinal Image Quality Using a Non-Mydriatic Fundus Camera in a Teleophthalmologic Platform

在远程眼科平台上使用非散瞳眼底照相机评估视网膜图像质量

阅读:1

Abstract

This study assesses the quality of retinal images captured using a non-mydriatic fundus camera within a teleophthalmologic platform in Taiwan. The objective was to evaluate the effectiveness of non-mydriatic fundus cameras for remote retinal screening and identify factors impacting image quality. From June 2020 to August 2022, 629 patients from five rural infirmaries underwent ophthalmic examinations, with fundus images captured without pupil dilation. These images were reviewed by senior ophthalmologists and graded based on quality. The results indicated that approximately 70% of images were of satisfactory diagnostic quality. Risk factors for poor image quality included older age, the presence of cataracts, pseudophakia, and diabetes mellitus. This study demonstrates the feasibility of using non-mydriatic fundus cameras for teleophthalmology, highlighting the importance of identifying and addressing factors that affect image quality to enhance diagnostic accuracy in remote settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。