Test-Retest Reproducibility of (18)F-FDG PET/CT Uptake in Cancer Patients Within a Qualified and Calibrated Local Network

在合格且经过校准的本地网络中,对癌症患者进行 (18)F-FDG PET/CT 摄取的测试-重测重复性分析

阅读:1

Abstract

Calibration and reproducibility of quantitative (18)F-FDG PET measures are essential for adopting integral (18)F-FDG PET/CT biomarkers and response measures in multicenter clinical trials. We implemented a multicenter qualification process using National Institute of Standards and Technology-traceable reference sources for scanners and dose calibrators, and similar patient and imaging protocols. We then assessed SUV in patient test-retest studies. Methods: Five (18)F-FDG PET/CT scanners from 4 institutions (2 in a National Cancer Institute-designated Comprehensive Cancer Center, 3 in a community-based network) were qualified for study use. Patients were scanned twice within 15 d, on the same scanner (n = 10); different but same model scanners within an institution (n = 2); or different model scanners at different institutions (n = 11). SUV(max) was recorded for lesions, and SUV(mean) for normal liver uptake. Linear mixed models with random intercept were fitted to evaluate test-retest differences in multiple lesions per patient and to estimate the concordance correlation coefficient. Bland-Altman plots and repeatability coefficients were also produced. Results: In total, 162 lesions (82 bone, 80 soft tissue) were assessed in patients with breast cancer (n = 17) or other cancers (n = 6). Repeat scans within the same institution, using the same scanner or 2 scanners of the same model, had an average difference in SUV(max) of 8% (95% confidence interval, 6%-10%). For test-retest on different scanners at different sites, the average difference in lesion SUV(max) was 18% (95% confidence interval, 13%-24%). Normal liver uptake (SUV(mean)) showed an average difference of 5% (95% confidence interval, 3%-10%) for the same scanner model or institution and 6% (95% confidence interval, 3%-11%) for different scanners from different institutions. Protocol adherence was good; the median difference in injection-to-acquisition time was 2 min (range, 0-11 min). Test-retest SUV(max) variability was not explained by available information on protocol deviations or patient or lesion characteristics. Conclusion:(18)F-FDG PET/CT scanner qualification and calibration can yield highly reproducible test-retest tumor SUV measurements. Our data support use of different qualified scanners of the same model for serial studies. Test-retest differences from different scanner models were greater; more resolution-dependent harmonization of scanner protocols and reconstruction algorithms may be capable of reducing these differences to values closer to same-scanner results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。