Comparison of the left ventricular dyssynchrony between stylet-driven and lumen-less lead technique in left bundle branch area pacing using myocardial perfusion scintigraphy

采用心肌灌注显像比较左束支区域起搏中,有导丝驱动导线和无腔导线技术引起的左心室不同步性

阅读:1

Abstract

BACKGROUND: Left bundle branch area pacing (LBBAP) has emerged as a novel physiological pacing method to reduce left ventricular (LV) dyssynchrony due to ventricular pacing. Only lumen-less pacing leads (LLLs) with fixed helixes could achieve LBBAP previously, but recently, LBBAP has been performed using stylet-driven leads (SDLs). This study aimed to evaluate the LV dyssynchrony between SDLs and LLLs techniques in LBBAP. METHODS: We retrospectively evaluated patients who underwent LBBAP with either SDLs or LLLs. We compared both groups' electrocardiogram (ECG) findings and LV dyssynchrony parameters derived from myocardial perfusion scintigraphy. LV dyssynchrony parameters consisted of phase analysis and regional wall motion analysis. We evaluated bandwidth, phase standard deviation (PSD), and entropy in the phase analysis. The time to the end-systolic frame (TES) was calculated in regional wall motion analysis using single-photon emission computed tomography (SPECT). We also evaluated the maximum differences between segmental TES (MDTES), the standard deviation of TES (SDTES), and the difference in the TES between the lateral wall and septum (DTES-LS). RESULTS: In total, 97 patients were enrolled. The success rate of LBBAP did not differ between the groups [SDLs: 47/48 patients (98%) vs. LLLs: 47/51 patients (92%), P=0.36]. The paced QRS duration and the stimulus to the peak LV activation time (stim-LVAT) also did not differ between SDL and LLL groups (122±10 vs. 119±12 ms, P=0.206; 69±12 vs. 66±13 ms, P=0.31, respectively). There were no differences in bandwidth, PSD, and entropy between SDL and LLL groups (73°±37° vs. 86°±47°, P=0.18; 19°±8.5° vs. 21°±9.7°, P=0.19; 0.57±0.08 vs. 0.59±0.08, P=0.17, respectively). The regional wall motion analysis parameters MDTES, SDTES, and DTES-LS also did not differ between SDL and LLL groups (19%±10% vs. 20%±10%, P=0.885; 5.0%±2.5% vs. 5.0%±2.5%, P=0.995; 5.0%±3.7% vs. 4.8%±4.2%, P=0.78, respectively). CONCLUSIONS: LBBAP using SDLs was comparable to LV electrical and mechanical synchrony with LLLs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。