Intraoperative Fracture during the Insertion of Advanced Locking Screws (T2 Alpha Femur Retrograde Intramedullary Nailing System): Report of Two Cases and Identifying Causes and Prevention

术中植入高级锁定螺钉(T2α股骨逆行髓内钉系统)时发生骨折:两例报告及原因分析与预防

阅读:1

Abstract

Background: Recently, the T2 alpha nailing system (Stryker, Inc.), which has advanced locking screws that can attach a screw to a rod, has been used worldwide and is expected to improve fracture fixation. We analyzed two cases of supracondylar femoral fractures in older adult patients, in which intraoperative fractures occurred during the insertion of advanced locking screws of the T2 alpha femur retrograde intramedullary nail. Case presentation: A 93-year-old and an 82-year-old woman each underwent T2 alpha femur retrograde nail fixation for supracondylar femur fractures at separate hospitals, and advanced locking screws were used as the proximal transverse locking screws. In both patients, a fracture line was observed at the proximal screw postoperatively, and the fractures were refixed with distal cable wiring and/or femoral distal plates. The patients were subsequently discharged from the same facility with no remarkable pain. Conclusions: When inserting advanced locking screws, it is necessary to enlarge the screw hole in the near-bone cortex with a counterbore drill, which might add torque to the bone cortex that could result in fractures. If the sleeve is distant from the bone, the counterbore drill will not reach the bone, the screw hole will not expand, and the insertion of advanced locking screws will apply a strong torque to the bone cortex and may result in fracture. Moreover, it is important to confirm that the counterbore drill is securely inserted under fluoroscopy and to carefully enlarge the bony foramen manually to prevent fractures during screw insertion.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。