Evaluation of Trabecular Microstructure of Cancellous Bone Using Quarter-Detector Computed Tomography

利用四分之一探测器计算机断层扫描评估松质骨小梁微结构

阅读:1

Abstract

Quarter-detector computed tomography (QDCT) is an ultra-high-spatial-resolution imaging technique. This study aimed to verify the validity of trabecular structure evaluation using a QDCT scanner in the diagnosis of osteoporosis. We used a cancellous bone specimen image of the second lumbar vertebrae of an adult male with moderate osteoporosis. To obtain QDCT images, we created a three-dimensional model from micro-CT images of the specimen. Statistical analysis was performed on the relationship between micro-CT and QDCT imaging modalities. The differences between micro-CT and QDCT were assessed based on their significance with respect to the calculated mean measurements using the Mann-Whitney test. Single regression analysis was performed using linear regression, with micro-CT and QDCT as the explanatory and objective variables, respectively, to determine the relationship of the measured values between the two modalities. By applying the necessary correction to the micro-CT measured values, it is possible to perform an analysis equivalent to micro-CT, which offers higher spatial resolution than QDCT. We found evidence that if QDCT can be used, trabecular structure evaluation may contribute to image diagnosis to evaluate practical bone fragility.