Validation of Optimum ROI Size for (123)I-FP-CIT SPECT Imaging Using a 3D Mathematical Cylinder Phantom

利用三维数学圆柱体模体验证 (123)I-FP-CIT SPECT 成像的最佳 ROI 尺寸

阅读:1

Abstract

OBJECTIVES: The partial volume effect (PVE) of single-photon emission computed tomography (SPECT) on corpus striatum imaging is caused by the underestimation of specific binding ratio (SBR). A large ROI (region of interest) set using the Southampton method is independent of PVE for SBR. The present study aimed to determine the optimal ROI size with contrast and SBR for striatum images and validate the Southampton method using a three-dimensional mathematical cylinder (3D-MAC) phantom. METHODS: We used ROIs sizes of 27, 36, 44, 51, 61, 68, and 76 mm for targets with diameters 40, 20, and 10 mm on reference and processed images reconstructed using the 3D-MAC phantom. Contrast values and SBR were compared with the theoretical values to obtain the optimal ROI size. RESULTS: The contrast values in the ROI with diameters of 51 (target: 40 mm in diameter) and 44 (target: 20 mm in diameter) mm matched the theoretical values. However, this value did not correspond with the 10-mm-diameter target. The SBR matched the theoretical value with an ROI of > 44 mm in the 20-mm-diameter target; but, it was under- and overestimated under any other conditions. CONCLUSION: These results suggested that an ROI should be 2-4 folds larger than the target size without PVE, and that the Southampton method was remarkably accurate.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。