The viscoelastic properties of the lung have important implications during respiratory mechanics in terms of lung movement or work of breathing, for example. However, this property has not been well characterized due to several reasons, such as the complex nature of the lung, difficulty accessing its tissues, and the lack of physical simulators that represent viscoelastic effects. This research proposes an electropneumatic system and a method to simulate the viscoelastic effect from temporary forces generated by the opposition of magnetic poles. The study was tested in a mechanical ventilation scenario with inspiratory pause, using a Hamilton-S1 mechanical ventilator (Hamilton Medical) and a simulator of the human respiratory system (SAMI-SII). The implemented system was able to simulate the stress relaxation response of a Standard Linear Solid model in the Maxwell form and showed the capacity to control elastic and viscous parameters independently. To the best of our knowledge, this is the first system incorporated into a physical lung simulator that represents the viscoelastic effect in a mechanical ventilation scenario.