Relationships between patient characteristics and contrast agent dose for successful computed tomography venography with a body-weight-tailored contrast protocol

采用体重定制对比剂方案,探讨患者特征与对比剂剂量对计算机断层扫描静脉造影成功率的影响

阅读:1

Abstract

The aim of this study was to evaluate the effect of patient characteristics on the contrast agent dosage that is required to reach effective enhancement of the inferior vena cava (IVC) on computed tomography venographs (CTV).This retrospective study included 50 patients who underwent CTV at 80 kVp. The contrast injection protocol (iodine 600 mg/kg) was tailored to their body weight. We calculated the required contrast agent volume (CAVmean-IVC) to reach the mean enhancement of IVC. We performed univariate and multivariate linear regression analyses between the sex, age, body weight (BW), lean body weight (LBW), body surface area (BSA), height (HT), estimated glomerular filtration rate (eGFR), and CAVmean-IVC.The univariate linear regression analysis show that HT, BW, LBW, and BSA were significantly correlated with CAVmean-IVC (P < .01 for all). The CAVmean-IVC was significantly higher for males than females (P < .01). Multivariate regression analysis showed that BW, LBW, and BSA had a statistically significant effect on CAVmean-IVC. There was no significant correlation of age, HT, or eGFR with CAVmean-IVC.BW, LBW, and BSA each had an independent significant effect on CAVmean-IVC. The conventional BW-tailored contrast injection protocol might be insufficient for CTV.