Evaluation of a one-step sample preparation protocol for analysis of total serum testosterone by LC-MS/MS

评估用于液相色谱-串联质谱法分析血清总睾酮的一步样品制备方案

阅读:1

Abstract

In this study, a one-step method for liquid-liquid extraction has been compared against a two-step procedure for testosterone assays in terms of accuracy, specificity, recovery, lipid removal and baseline noise, using QCs and unknown samples. The difference in accuracy was less than 5% for adult sera, while it was less than 10% for prepubescent sera. To compare specificity, the ion ratio transition of 289 → 97 to 289 → 109 was monitored for all QCs and unknown samples; no interference in the testosterone peak was observed for any tested sample prepared by either the one-step or two-step procedure. The baseline comparison of LC-MS/MS chromatograms of samples indicated that samples prepared by the one-step procedure were of the same quality as those prepared by the two-step procedure; however, recovery in unaltered serum using the one-step procedure was approximately 15% greater across the low to high concentration range. Furthermore, recovery using the one-step procedure was more consistent between stripped and unstripped serum. Lipids were removed efficiently in the two-step procedure as verified by monitoring the typical phospholipid MRM transition of m/z 496 → 184. For every sample processed by the one-step procedure, a one-minute online column wash with 95% methanol was able to remove 95% of the bound lipids, thereby providing a column life-time approximately equivalent to that for the two-step procedure. The presented data indicate that the one-step procedure could replace the two-step procedure while maintaining accuracy, saving time, increasing recovery, and minimizing the potential for errors with the fewer steps required.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。