Fracture strength of orthodontic mini-implants

正畸微型植入物的断裂强度

阅读:7
作者:Tatiana Feres Assad-Loss, Flávia Mitiko Fernandes Kitahara-Céia, Giordani Santos Silveira, Carlos Nelson Elias, José Nelson Mucha

Conclusions

: The fracture torques were different for both the neck and the tip of the five types evaluated. NEO and SIN mini-implants showed the highest resistance to fracture of the neck and tip. The fracture torques of both tip and neck were higher than the torque required to insert mini-implants. Objetivo:: o objetivo do presente estudo foi avaliar o desenho e as medidas de cinco marcas diferentes de mini-implantes ortodônticos, e sua influência na resistência à fratura em torção. Métodos:: cinquenta mini-implantes foram divididos em cinco grupos, correspondentes a diferentes fabricantes (DEN, Dentaurum; RMO, Rocky Mountain Orthodontics; CON, Conexão; NEO, Neodent; SIN, Sistema de Implantes Nacional). Vinte e cinco mini-implantes foram submetidos ao teste de fratura por torção no pescoço e na ponta, com mandris fixados a uma máquina universal de testes mecânicos. Os outros 25 mini-implantes foram submetidos ao teste de torque de inserção em blocos de costelas suínas, utilizando-se um torquímetro e um contra-ângulo montado em motor cirúrgico. O formato da ponta ativa dos mini-implantes foi avaliado por microscopia. O teste não-paramétrico de Friedman e o teste F de Snedecor na análise de variância (ANOVA) foram utilizados para avaliar as diferenças entre os grupos. Resultados:: o torque de fratura do pescoço variou de 23,45 N.cm (DEN) a 34,82 N.cm (SIN); e o da ponta, entre 9,35 N.cm (CON) e 24,36 N.cm (NEO). O valor do torque de inserção variou de 6,6 N.cm (RMO) a 10,2 N.cm (NEO). As características que mais influenciaram os resultados foram: diâmetro externo, diâmetro interno, razão entre o diâmetro interno e o externo, e a presença de fresagem na região apical do mini-implante. Conclusões:: os torques de fratura foram diferentes entre os cinco tipos avaliados, tanto no pescoço quanto na ponta. Os mini-implantes NEO e SIN foram os mais resistentes à fratura do pescoço e da ponta. Os torques de fratura tanto na ponta quanto no pescoço foram maiores do que o torque necessário para a inserção dos mini-implantes.

Methods

: Fifty mini-implants were divided into five groups corresponding to different manufactures (DEN, RMO, CON, NEO, SIN). Twenty-five mini-implants were subjected to fracture test by torsion in the neck and the tip, through arbors attached to a Universal Mechanical Testing Machine. The other 25 mini-implants were subjected to insertion torque test into blocks of pork ribs using a torquimeter and contra-angle handpiece mounted in a surgical motor. The shape of the active tip of the mini-implants was evaluated under microscopy. The non-parametric Friedman test and Snedecor's F in analysis of variance (ANOVA) were used to evaluate the differences between groups.

Objective

: This study aimed at evaluating the design and dimensions of five different brands of orthodontic mini-implants, as well as their influence on torsional fracture strength.

Results

: The fracture torque of the neck ranged from 23.45 N.cm (DEN) to 34.82 N.cm (SIN), and of the tip ranged from 9.35 N.cm (CON) to 24.36 N.cm (NEO). Insertion torque values ranged from 6.6 N.cm (RMO) to 10.2 N.cm (NEO). The characteristics that most influenced the results were outer diameter, inner diameter, the ratio between internal and external diameters, and the existence of milling in the apical region of the mini-implant. Conclusions:: The fracture torques were different for both the neck and the tip of the five types evaluated. NEO and SIN mini-implants showed the highest resistance to fracture of the neck and tip. The fracture torques of both tip and neck were higher than the torque required to insert mini-implants. Objetivo:: o objetivo do presente estudo foi avaliar o desenho e as medidas de cinco marcas diferentes de mini-implantes ortodônticos, e sua influência na resistência à fratura em torção. Métodos:: cinquenta mini-implantes foram divididos em cinco grupos, correspondentes a diferentes fabricantes (DEN, Dentaurum; RMO, Rocky Mountain Orthodontics; CON, Conexão; NEO, Neodent; SIN, Sistema de Implantes Nacional). Vinte e cinco mini-implantes foram submetidos ao teste de fratura por torção no pescoço e na ponta, com mandris fixados a uma máquina universal de testes mecânicos. Os outros 25 mini-implantes foram submetidos ao teste de torque de inserção em blocos de costelas suínas, utilizando-se um torquímetro e um contra-ângulo montado em motor cirúrgico. O formato da ponta ativa dos mini-implantes foi avaliado por microscopia. O teste não-paramétrico de Friedman e o teste F de Snedecor na análise de variância (ANOVA) foram utilizados para avaliar as diferenças entre os grupos. Resultados:: o torque de fratura do pescoço variou de 23,45 N.cm (DEN) a 34,82 N.cm (SIN); e o da ponta, entre 9,35 N.cm (CON) e 24,36 N.cm (NEO). O valor do torque de inserção variou de 6,6 N.cm (RMO) a 10,2 N.cm (NEO). As características que mais influenciaram os resultados foram: diâmetro externo, diâmetro interno, razão entre o diâmetro interno e o externo, e a presença de fresagem na região apical do mini-implante. Conclusões:: os torques de fratura foram diferentes entre os cinco tipos avaliados, tanto no pescoço quanto na ponta. Os mini-implantes NEO e SIN foram os mais resistentes à fratura do pescoço e da ponta. Os torques de fratura tanto na ponta quanto no pescoço foram maiores do que o torque necessário para a inserção dos mini-implantes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。