Changes in the specific absorption rate (SAR) of radiofrequency energy in patients with retained cardiac leads during MRI at 1.5T and 3T

在1.5T和3T磁共振成像中,保留心脏导线的患者射频能量比吸收率(SAR)的变化

阅读:1

Abstract

PURPOSE: To evaluate the local specific absorption rate (SAR) and heating around retained cardiac leads during MRI at 64 MHz (1.5T) and 127 MHz (3T) as a function of RF coil type and imaging landmark. METHODS: Numerical models of retained cardiac leads were built from CT and X-ray images of 6 patients with retained cardiac leads. Electromagnetic simulations and bio-heat modeling were performed with MRI RF body and head coils tuned to 64 MHz and 127 MHz and positioned at 9 different imaging landmarks covering an area from the head to the lower limbs. RESULTS: For all patients and at both 1.5T and 3T, local transmit head coils produced negligible temperature rise ( ΔT < 0.1°C ) for ‖‖B1+‖‖ ≤ 3 μT . For body imaging with quadrature-driven coils at 1.5T, ΔT during a 10-min scan remained < 3°C at all imaging landmarks for ‖‖B1+‖‖ ≤ 3 μT and <6°C for ‖‖B1+‖‖ ≤ 4 μT . For body imaging at 3T, ΔT during a 10-min scan remained < 6°C at all imaging landmarks for ‖‖B1+‖‖ ≤ 2 μT . For shorter pulse sequences up to 2 min, ΔT remained < 6°C for ‖‖B1+‖‖ ≤ 3 μT . CONCLUSION: For the models based on 6 patients studied, simulations suggest that MRI could be performed safely using a local head coil at both 1.5T and 3T, and with a body coil at 1.5T with pulses that produced ‖‖B1+‖‖ ≤ 4 μT . MRI at 3T could be performed safely in these patients using pulses with ‖‖B1+‖‖ ≤ 2 μT .