Performance Evaluation of Open Channel Buhlmann Fecal Calprotectin Turbo Assay on Abbott Alinity C Analyzer

Abbott Alinity C 分析仪上开放通道 Buhlmann 粪便钙卫蛋白快速检测法的性能评估

阅读:1

Abstract

Inflammatory bowel disease (IBD) is characterized by chronic inflammation of the gastrointestinal (GI) tract. Fecal calprotectin (fCAL) is a noninvasive laboratory test used in the diagnosis and monitoring of IBDs such as Crohn's disease and ulcerative colitis. The fCAL send-out test that our facility has been offering so far uses an ELISA-based method. In the current study, we sought to validate the performance of a Buhlmann fCAL turbo assay in an automated Abbott Alinity C analyzer (AFCAL) in our core laboratory. Five-day imprecision studies showed good performance for both within-run (5.3%) and between-day (2.5%) measurements. The reportable range was verified as 30-20,000 µg/g. Deming regression and Bland-Altman analysis indicated a strong correlation of r = 0.99 with a low, acceptable bias of 1.8% for AFCAL relative to the predicate Buhlmann fCAL ELISA results. AFCAL's clinical performance was determined retrospectively in 62 patients with ICD codes for IBD. Overall, the implementation of AFCAL in our routine clinical testing has improved our turnaround time, reduced the cost per test, and significantly increased our clinician satisfaction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。