Assessing the elastic properties of skeletal muscle and tendon using shearwave ultrasound elastography and MyotonPRO

利用剪切波超声弹性成像和MyotonPRO评估骨骼肌和肌腱的弹性特性

阅读:1

Abstract

The purposes of this study were to compare Young's modulus values determined by shear wave ultrasound elastography (SWUE) with stiffness index obtained using a hand-held MyotonPRO device on the resting stiffness of gastrocnemius muscle belly and Achilles tendon; and to examine the test-retest reliability of those stiffness measurement using hand-held MyotonPRO. Twenty healthy volunteers participated in the study. The measurement values of muscle and tendon was determined in dominant legs. Each marker point was assessed using MyotonPRO and SWUE, respectively. Intra-operator reliability of MyotonPRO was established in 10 of the subjects. The correlation coefficients between the values of muscle and tendon stiffness indices determined by MyotonPRO and SWUE were calculated. Significant correlations were found for muscle and tendon stiffness and Young's modulus ranged from 0.463 to 0.544 (all P < 0.05). The intra-operator reliability ranged from good to excellent (ICC((3,1)) = 0.787~0.928). These results suggest that the resting stiffness of gastrocnemius muscle belly and Achilles tendon measured by MyotonPRO is related to the Young's modulus of those quantified by SWUE. The MyotonPRO shows good intra-operator repeatability. Therefore, the present study shows that MyotonPRO can be used to assess mechanical properties of gastrocnemius muscle belly and Achilles tendon with a resting condition.