Assessing the cyclic fatigue resistance and sterilization effects on replica-like endodontic instruments compared to Reciproc Blue

评估复制型根管器械与 Reciproc Blue 相比的循环疲劳强度和消毒效果

阅读:1

Abstract

This study aimed to evaluate the effect of the number of uses and autoclave sterilization on the cyclic fatigue resistance of four replica-like instruments RC Blue; Only One File Blue; Recip One Blue; and Micro Blue compared to the original system Reciproc Blue (VDW, Munich, Germany). The instruments were analyzed by scanning electron microscope (SEM) before being used in root canal instrumentation (baseline). Fifty molars were instrumented by ten instruments (n=10). After sterilization in an autoclave, the  instruments were analyzed by SEM. This procedure was repeated twice more using different molars, totaling 3 rounds of instrumentation, sterilization and SEM analysis. Then, ten different instruments from each brand were tested for cyclic fatigue resistance. Number of uses data were analyzed using Chi-squared analysis, and cyclic fatigue data were analyzed by one-way Anova followed by Tukey's test as the data had normal distribution. The fracture times for all systems had no significant difference, but Micro Blue had significantly lower values than the other systems (p < 0.05). The SEM analysis showed distortions in the instruments after the 3rd use. Therefore, all tested instruments except of Micro Blue have similar resistance to cyclic fatigue, and all are reliable for use in up to 2-cases.