[Research on a portable shielding-free ultra-low field magnetic resonance imaging system]

【便携式无屏蔽超低场磁共振成像系统的研究】

阅读:1

Abstract

The portable light-weight magnetic resonance imaging system can be deployed in special occasions such as Intensive Care Unit (ICU) and ambulances, making it possible to implement bedside monitoring imaging systems, mobile stroke units and magnetic resonance platforms in remote areas. Compared with medium and high field imaging systems, ultra-low-field magnetic resonance imaging equipment utilizes light-weight permanent magnets, which are compact and easy to move. However, the image quality is highly susceptible to external electromagnetic interference without a shielded room and there are still many key technical problems in hardware design to be solved. In this paper, the system hardware design and environmental electromagnetic interference elimination algorithm were studied. Consequently, some research results were obtained and a prototype of portable shielding-free 50 mT magnetic resonance imaging system was built. The light-weight magnet and its uniformity, coil system and noise elimination algorithm and human brain imaging were verified. Finally, high-quality images of the healthy human brain were obtained. The results of this study would provide reference for the development and application of ultra-low-field magnetic resonance imaging technology.