Enhancing Veress Needle Entry with Proximal Vibroacoustic Sensing for Automatic Identification of Peritoneum Puncture

利用近端振动声学传感增强Veress针穿刺,实现腹膜穿刺的自动识别

阅读:1

Abstract

Laparoscopic access, a critical yet challenging step in surgical procedures, often leads to complications. Existing systems, such as improved Veress needles and optical trocars, offer limited safety benefits but come with elevated costs. In this study, a prototype of a novel technology for guiding needle interventions based on vibroacoustic signals is evaluated in porcine cadavers. The prototype consistently detected successful abdominal cavity entry in 100% of cases during 193 insertions across eight porcine cadavers. The high signal quality allowed for the precise identification of all Veress needle insertion phases, including peritoneum puncture. The findings suggest that this vibroacoustic-based guidance technology could enhance surgeons' situational awareness and provide valuable support during laparoscopic access. Unlike existing solutions, this technology does not require sensing elements in the instrument's tip and remains compatible with medical instruments from various manufacturers.