The Effects of Novel Pulsed Electromagnetic Field Therapy Device on Acute Distal Radius Fractures: A Prospective, Double-Blind, Sham-Controlled, Randomized Pilot Study

新型脉冲电磁场治疗装置对急性桡骨远端骨折的影响:一项前瞻性、双盲、假手术对照、随机试点研究

阅读:1

Abstract

BACKGROUND: this pilot study aimed at determining whether the application of a novel new method of generating pulsed electromagnetic field (PEMF), the Fracture Healing Patch (FHP), accelerates the healing of acute distal radius fractures (DRF) when compared to a sham treatment. METHODS: 41 patients with DRFs treated with cast immobilization were included. Patients were allocated to a PEMF group (n = 20) or a control (sham) group (n = 21). All patients were assessed with regard to functional and radiological outcomes (X-rays and CT scans) at 2, 4, 6 and 12 weeks. RESULTS: fractures treated with active PEMF demonstrated significantly higher extent of union at 4 weeks as assessed by CT (76% vs. 58%, p = 0.02). SF12 mean physical score was significantly higher in PEMF treated group (47 vs. 36, p = 0.005). Time to cast removal was significantly shorter in PEMF treated patients, 33 ± 5.9 days in PEMF vs. 39.8 ± 7.4 days in sham group (p = 0.002). CONCLUSION: early addition of PEMF treatment may accelerate bone healing which could lead to a shorter cast immobilization, thus allowing an earlier return to daily life activities and work. There were no complications related to the PEMF device (FHP).