Real time visualization and characterization of platelet deposition under flow onto clinically relevant opaque surfaces

实时可视化和表征流动状态下血小板在临床相关的不透明表面上的沉积情况

阅读:14
作者:Megan A Jamiolkowski, Joshua R Woolley, Marina V Kameneva, James F Antaki, William R Wagner

Abstract

Although the thrombogenic nature of the surfaces of cardiovascular devices is an important aspect of blood biocompatibility, few studies have examined platelet deposition onto opaque materials used for these devices in real time. This is particularly true for the metallic surfaces used in current ventricular assist devices (VADs). Using hemoglobin depleted red blood cells (RBC ghosts) and long working distance optics to visualize platelet deposition, we sought to perform such an evaluation. Fluorescently labeled platelets mixed with human RBC ghosts were perfused across six opaque materials (a titanium alloy (Ti6Al4V), silicon carbide (SiC), alumina (Al2O3, 2-methacryloyloxyethyl phosphorylcholine polymer coated Ti6Al4V (MPC-Ti6Al4V), yttria partially stabilized zirconia (YZTP), and zirconia toughened alumina (ZTA)) for 5 min at wall shear rates of 400 and 1000 s(-1). Ti6Al4V had significantly increased platelet deposition relative to MPC-Ti6Al4V, Al2 O3 , YZTP, and ZTA at both wall shear rates (p < 0.01). For all test surfaces, increasing the wall shear rate produced a trend of decreased platelet adhesion. The described system can be a utilized as a tool for comparative analysis of candidate blood-contacting materials with acute blood contact.