Impact of Reference Selection on Gamma Analysis in Patient-Specific Quality Assurance for Monte Carlo-Based Treatment Planning Systems

参考值选择对基于蒙特卡罗方法的治疗计划系统中患者特定质量保证的伽马分析的影响

阅读:1

Abstract

OBJECTIVE: Our study aimed to establish a standardized methodology for selecting "reference" and "evaluated" distributions in gamma analysis for Monte Carlo (MC) based intensity modulated treatment plans. Evaluation of importance of reference selection in MC based and non MC based treatment planning systems were analysed using a study classification. METHODS: Three categories were utilized to analyzed gamma passing rates across using different treatment planning systems (TPS) and detectors for thirty five patients. Category 1 utilized MC-based Monaco TPS plans and a 2 dimensional(2D) I'mRTMatriXX detector. Category 2 employed non-MC-based Eclipse TPS plans, assessed with a 2D I'mRTMatriXX detector. In Category 3, MC-based Monaco TPS plans were validated using a Dolphin detector. All plans were subjected to analysis using gamma criteria, which considered a dose difference of 3% and a distance to agreement of 3mm. Additionally, another set of gamma criteria was employed, with a dose difference of 3% and a distance to agreement of 2mm. An introduced Asymmetric factors in both 2D and 3D analysis will quantify the asymmetric nature of gamma based on the choice of "reference" distribution. RESULT: For 2D Gamma analysis, MC-based Monaco TPS and I'mRTMatriXX showed a consistent positive Zk2D trend for all patients, with significant p-values below 0.01 for both gamma passing criteria. In contrast, non-MC based Eclipse TPS exhibited varied Zk2D results, with non-significant p-values. In 3D Gamma analysis, all patients exhibited positive Zk3D values with significant p-values below 0.01 when "references" were swapped. The Pearson correlation between asymmetricity and isodose volumes was notably high at 0.99 for both gamma criteria. CONCLUSION: Our study highlights the imperative of using MC-based TPS as the definitive "reference" in gamma analysis for patient specific quality assurance of intensity modulated radiation therapy, emphasizing that variations can mislead results, especially given gamma analysis's sensitivity to MC calculation noise.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。