[Study on direct ventricular assist loading mode based on a finite element method]

[基于有限元法的直接心室辅助负荷模式研究]

阅读:1

Abstract

To investigate the biomechanical effects of direct ventricular assistance and explore the optimal loading mode, this study established a left ventricular model of heart failure patients based on the finite element method. It proposed a loading mode that maintains peak pressure compression, and compared it with the traditional sinusoidal loading mode from both hemodynamic and biomechanical perspectives. The results showed that both modes significantly improved hemodynamic parameters, with ejection fraction increased from a baseline of 29.33% to 37.32% and 37.77%, respectively, while peak pressure, stroke volume, and stroke work parameters also increased. Additionally, both modes showed improvements in stress concentration and excessive fiber strain. Moreover, considering the phase error of the assist device's working cycle, the proposed assist mode in this study was less affected. Therefore, this research may provide theoretical support for the design and optimization of direct ventricular assist devices.