A proof of principle study using radiopharmaceuticals to quantify and localize container-content interactions in medical syringes

利用放射性药物量化和定位医用注射器中容器-内容物相互作用的原理验证研究

阅读:1

Abstract

The sorption of drugs onto their contents is a known phenomenon that is difficult to analyse precisely. The purpose of this study was to present a non-invasive method for locating and quantifying sorption phenomena using radiopharmaceuticals. Radiopharmaceutical are medicines armed with a radionuclide enabling quantification and imaging using dedicated scanners. The sorption of nine different radiopharmaceuticals on 2- and 3-part syringes was investigated. These syringes were filled with the studied radiopharmaceutical solutions and stored immobile for 3 h. At different times ranging from 0 to 180 min, 10 µL were taken from the syringes and the radioactivity of these samples was determined by a gamma counter. 5 radiopharmaceuticals exhibited no significant sorption at any time point in both 2 and 3-parts syringes, but 4 radiopharmaceuticals exhibited sorption losses varying from 20 to 33% after 3 h contact with 3-part-syringes, but no sorption on 2-part syringes at any time point. [(99m)Tc]Tc-tetrofosmine Single Photon Emission Computed Tomography/Computed Tomography imaging indicated clearly that the interactions were located on the rubber plunger of the 3-part-syringes. The specific nature of radiopharmaceuticals allowed their use as an innovative method to quantify and localize drug sorption phenomena.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。