Ultrasonographic reliability and repeatability of simultaneous bilateral assessment of diaphragm muscle thickness during normal breathing

正常呼吸过程中同时双侧评估膈肌厚度的超声检查可靠性和重复性

阅读:1

Abstract

BACKGROUND: The diaphragm is considered the main muscle involved in breathing and also linked to trunk stabilization functions. Up to date, rehabilitative ultrasound imaging (RUSI) has been the most used technique to evaluate unilaterally the transcostal diaphragm thickness. Nevertheless, the inspiratory activity of both hemi-diaphragms is bilaterally performed at the same time, and its simultaneous evaluation with a thoracic orthosis could improve its assessment as well as its re-education with visual biofeedback of both hemi-diaphragms at the same time. The purpose was to evaluate the reliability and repeatability of simultaneous thickness measurements of both hemi-diaphragms bilaterally during normal breathing using a thoracic orthosis that allowed bilateral fixation of both right and left ultrasound probes. METHODS: The study was conducted in 46 healthy subjects, whose diaphragm thickness was measured bilaterally and simultaneously in the anterior axillary line during relaxed breathing with a designed thoracic orthosis and 2 ultrasound tools. Intra-examiner (same examiner), inter-examiner (2 examiners), intra-session (1 hour) and inter-session (1 week) reliability and repeatability between each pair of measurements of diaphragm muscle thickness were analyzed during normal breathing. RESULTS: Reliability and repeatability for intra-session evaluations using the thoracic orthosis were excellent to evaluate simultaneous thickness of both hemi-diaphragms by bilateral probes fixation (intraclass correlation coefficient =0.919-0.997; standard error of measurement =0.002-0.007 cm; minimum detectable change =0.006-0.020 cm), without systematic errors (P>0.05) between each pair of measurements. Nevertheless, inter-session evaluations varied from good to excellent using the bilateral probes fixation (intraclass correlation coefficient =0.614-0.984; standard error of measurement =0.006-0.028 cm; minimum detectable change =0.017-0.079 cm), although some systematic errors were presented (P<0.05). CONCLUSIONS: Good to excellent reliability and repeatability was shown for simultaneous thickness measurements of both hemi-diaphragms bilaterally during normal breathing. Despite systematic errors were presented for some inter-examiner assessments, the use of the thoracic orthosis that allowed bilateral fixation of ultrasound probes could be recommended for simultaneous hemi-diaphragms breathing re-education by visual biofeedback.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。