Exploration of quality criteria for the detection of peak inspiratory flow under different resistance conditions

探索不同阻力条件下峰值吸气流速检测的质量标准

阅读:1

Abstract

BACKGROUND: At present, robust quality criteria and methods for the assessment of Peak inspiratory flow meter performance are lacking. OBJECTIVE: A standard flow-volume simulator for quality control analyses of an inhalation assessment device was utilized with different simulated resistance levels in order to propose a quality testing method and associated standard for this device type. METHODS: A standard flow-volume simulator was utilized to assess the performance of an In-Check DIAL® (Device I) and an intelligent inhalation assessment device (Device P) at a fixed volume and flow rate. Indices used to evaluate these two instruments included repeatability, accuracy, linearity, and impedance. RESULTS: Both devices exhibited good repeatability (<± 3 L/min). The difference between test results and standard simulator values for Device P was less than ± 5 L/min at resistance level R1 but higher than ± 5 L/min at resistance levels R2-5, while Device I were greater than 5 L/min at all resistance levels. The relative error for Device P was <± 10% at resistance levels R1, R2, and R4, but > 10% at resistance levels R3 and R5. The relative error values for Device I at all five resistance levels were > 10%. Device P passed the linearity test at the R2 resistance level, while Device I partially passed the linearity test at all five resistance levels. CONCLUSION: Standard monitoring methods and standards provide a valuable approach to the more reliable clinical assessment and application of these instruments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。