Successful Retrieval of Rota Burr After Driveshaft Fracture

驱动轴断裂后成功取出旋转毛刺

阅读:1

Abstract

Rotational atherectomy is an effective procedure for heavily calcified lesions and those that cannot be crossed using conventional percutaneous coronary intervention (PCI) devices. Here, we report a rare case of intracoronary burr entrapment in the coronary artery due to burr disconnection from the driveshaft. A 67-year-old man undergoing hemodialysis for nephrosclerosis presented with exertional chest discomfort. Coronary angiography revealed stenotic lesions in the right coronary artery, and PCI was performed using a Rotawire Floppy. During the procedure, the disconnected burr was successfully removed without surgery using the child-in-mother technique with a guide extension catheter. Notably, the patient remained hemodynamically stable throughout the procedure and his recovery was uncomplicated. He was discharged on the second postprocedural day. At the 6-month follow-up, the patient remained asymptomatic with no evidence of myocardial ischemia. This report informs clinicians of the possibility of burr disconnection and the non-surgical intervention used for its removal.