A Novel Locking Buttress Plate Designed for Simultaneous Medial and Posterolateral Tibial Plateau Fractures: Concept and Comparative Finite Element Analysis

一种用于同时治疗内侧和后外侧胫骨平台骨折的新型锁定支撑钢板:概念和比较有限元分析

阅读:1

Abstract

OBJECTIVE: The complex tibial plateau fractures involving both medial and posterolateral columns are of frequent occurrence in clinics, but the existing fixation system cannot deal with medial and posterolateral fragments simultaneously. Therefore, a novel locking buttress plate named as medial and posterior column plate (MPCP) was designed in this study to fix the simultaneous medial and posterolateral tibial plateau fractures. Meanwhile, the comparative finite element analysis (FEA) was conducted to investigate the discrepancy between MPCP and traditional multiple plates (MP + PLP) in their biomechanical characteristics. METHODS: Two 3D finite element models of simultaneous medial and posterolateral tibial plateau fracture fixed with MPCP and MP + PLP system, respectively, was constructed. To imitate the axial stress of knee joint in ordinary life, diverse axial forces with 100, 500, 1000, and 1500 N were applied in the two fixation models, and then the equivalent displacement and stress nephograms and values were obtained. RESULTS: The similar trend of displacement and stress increasing with the loads was observed in the two fixation models. However, several heterogeneities of displacement and stress distribution were found in the two fixation models. The max displacement and von Mises stress values of plates, screws, and fragments in the MPCP fixation model were significantly smaller than that in the MP + PLP fixation model, except for the max-shear stress values. CONCLUSION: As a single locking buttress plate, the MPCP system showed the excellent benefit on improving the stability of the simultaneous medial and posterolateral tibial plateau fractures, compared with the traditional double plate fixation system. However, the excessive shear stress around screw holes should be paid attention to prevent trabecular microfracture and screw loosening.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。