Impact of Evolution of Self-Expandable Aortic Valve Design: Peri-Operative and Short-Term Outcomes

自膨式主动脉瓣膜设计演变的影响:围手术期和短期结果

阅读:1

Abstract

A few data exist on the differences of implantable aortic valve bio-prostheses. We investigate three generations of self-expandable aortic valves in terms of the outcomes. Patients undergoing transcatheter aortic valve implantation (TAVI) were allocated into three groups according to the valve type: group A (CoreValve(TM)), group B (Evolut(TM)R) and group C (Evolut(TM)PRO). The implantation depth, device success, electrocardiographic parameters, need for permanent pacemaker (PPM), and paravalvular leak (PVL) were assessed. In the study, 129 patients were included. The final implantation depth did not differ among the groups (p = 0.07). CoreValve(TM) presented greater upward jump of the valve at release (2.88 ± 2.33 mm vs. 1.48 ± 1.09 mm and 1.71 ± 1.35 mm, for groups A, B, and C, respectively, p = 0.011). The device success (at least 98% for all groups, p = 1.00) and PVL rates (67% vs. 58%, vs. 60% for groups A, B, and C, respectively, p = 0.64) did not differ. PPM implantation within 24 h (33% vs. 19% vs. 7% for groups A, B, and C, respectively, p = 0.006) and until discharge (group A: 38% vs. group B: 19% and group C: 9%, p = 0.005) was lower in the newer generation valves. Newer generation valves present better device positioning, more predictable deployment, and fewer rates of PPM implantation. No significant difference in PVL was observed.