RP-HPLC method validation for fast extraction and quantification of Levonorgestrel drug from silicone based intrauterine device intended for in-process and finished formulation

反相高效液相色谱法验证,用于快速提取和定量分析用于在制品和成品配方的硅胶宫内节育器中的左炔诺孕酮药物

阅读:5
作者:Midhu George Veeran, Karthikeyan C, Bharaniraja B, Diksha Painuly, Abi Santhosh Aprem

Background

To develop and validate a simple and consistent reversed phase high performance liquid chromatography (RP-HPLC) method for the estimation of Levonorgestrel (LNG) drug from silicone based intrauterine device.

Conclusion

The validated method is found to be a simple, accurate, precise, reproducible, and hence can be used for the routine analysis of LNG such as in-process, quality control and stability assays of silicone based intrauterine devices by RP-HPLC.

Methods

Sample solution was prepared using tetrahydrofuran (THF) as solvent for the drug extraction, and RP-HPLC analysis was performed using Luna C18 analytical column (150 × 4.6 mm, 5 μm, 100 Å - Phenomenex), with a mobile phase consisting of a mixture of acetonitrile and water (50:50, v/v) at a flow rate of 1.0 ml/min and injection volume of 20 μl. Detection was carried out at 241 nm in PDA detector, with a total run time of 15 min. The method was validated in accordance with ICH guidelines. Method applicability was tested for optimizing formulation using quality-by-design approach, to check the stability and content uniformity of levonorgestrel-silicone mixture (core blend), and quantifying the amount of LNG from commercially available silicone based formulation.

Results

The retention time for LNG drug was obtained at 8.5 min (± 0.3 min). A linear relationship was observed over the concentration range of 2.6-15.6 μg/ml with the correlation coefficient (r) value 0.9999. The method was found to be precise within the acceptable limit (RSD < 2%) and the drug recovery from the intrauterine device was found in the range 99.78-100.0%. Content uniformity for different prototypes developed was observed in the range of 91.6-101.4%, and assay of optimized core blend was in the range of 97.78-106.79% during the 10 days of retention period for stability studies.