A Novel and Reliable Pixel Response Correction Method (DAC-Shifting) for Spectral Photon-Counting CT Imaging

一种用于光谱光子计数CT成像的新型可靠像素响应校正方法(DAC偏移)

阅读:1

Abstract

Spectral photon-counting cone-beam computed tomography (CT) imaging is challenged by individual pixel response behaviours, which lead to noisy projection images and subsequent image artefacts like rings. Existing methods to correct for this either use calibration measurements, like signal-to-thickness calibration (STC), or perform a post-processing ring artefact correction of sinogram data or scan reconstructions without taking the pixel response explicitly into account. Here, we present a novel post-processing method (digital-to-analogue converter (DAC)-shifting) which explicitly measures the current pixel response using flat-field images and subsequently corrects the projection data. The DAC-shifting method was evaluated using a repeat series of the spectral photon-counting imaging (Medipix3) of a phantom with different density inserts and iodine K-edge imaging. The method was also compared against polymethyl methacrylate (PMMA)-based STC. The DAC-shifting method was shown to be effective in correcting individual pixel responses and was robust against detector instability; it led to a 47.4% average reduction in CT-number variation in homogeneous materials, with a range of 40.7-55.6%. On the contrary, the STC correction showed varying results; a 13.7% average reduction in CT-number variation, ranging from a 43.7% increase to a 45.5% reduction. In K-edge imaging, DAC-shifting provides a sharper attenuation peak and more uniform CT values, which are expected to benefit iodine concentration quantifications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。