Three-dimensional finite element analysis of stress distribution on different complex macro designs in commercially available implants: An in-vitro study

市售植入物不同复杂宏观设计应力分布的三维有限元分析:一项体外研究

阅读:1

Abstract

OBJECTIVE: This study aimed to investigate the effects of different commercially available complex implant macro designs on stress distributions using Finite element analysis. The experiment is done under varying simulated bone conditions to provide reference for clinical application. MATERIALS AND METHODS: The study employed the Finite Element Analysis (FEA) method to compare four commercially available complex implant macro designs on a Computer-Aided Design (CAD) model of a maxillary bone segment. The three-dimensional geometrical model of the implants was reconstructed from computed tomography (CT)-slices in Digital Imaging and Communications in Medicine (DICOM) format and contact condition between the implant and the bone was considered as 'Bonded', implying perfect osseointegration. All materials used in the models were assumed to be isotropic, homogeneous, and linearly elastic. The Finite element simulations employed load of 400 N under both axial and non-axial conditions Stresses were analysed under different bone conditions. RESULTS: Average values of von Mises stresses were used for comparing stress levels between implant designs. There was a definite increase in the equivalent stress values from higher density(D1)to lower density (D4) bone conditions for all implants. The percentage of increase ranged from 23.63 to 49.39 on axial loading and 20.39 to 57.19 when subjected to non-axial loading. The equivalent stress values resulted from non-axial loading were 1.78-2.94 times higher than that of axial loading for all implants under all bone densities. Among the complex designs Equinox Myriad Plus implant exhibited the least stress under axial loading (12.749-19.046 MPa) and (37.462-49.217 MPa) for non-axial loading. The stress on the crestal module was higher (1.49-2.99 times) than the overall stress on the implant regardless of the loading direction or bone conditions. CONCLUSIONS: Data from the present study shows Equinox Myriad Plus implant generating the least equivalent stress and this can be taken as indicator in the biomechanical performance of the design.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。