Diagnostic Performance of Point-of-Care Apparent Diffusion Coefficient Measures to Reduce Biopsy in Breast Lesions at MRI: Clinical Validation

床旁表观扩散系数测量在减少乳腺病灶MRI活检中的诊断性能:临床验证

阅读:1

Abstract

Background The Eastern Cooperative Oncology Group-American College of Radiology Imaging Network Cancer Research Group multicenter A6702 trial identified an optimal apparent diffusion coefficient (ADC) cutoff to potentially reduce biopsies by 21% without affecting sensitivity. Whether this performance can be achieved in clinical settings has not yet been established. Purpose To validate the performance of point-of-care ADC measurements with the A6702 trial ADC cutoff for reducing unnecessary biopsies in lesions detected at breast MRI. Materials and Methods Consecutive breast MRI examinations performed from May 2015 to January 2019 at a single medical center and showing biopsy-confirmed Breast Imaging Reporting and Data System category 4 or 5 lesions, without ipsilateral cancer, were identified. Point-of-care lesion ADC measurements collected at clinical interpretation were retrospectively evaluated. MRI examinations included axial T2-weighted, diffusion-weighted, and dynamic contrast-enhanced sequences. Sensitivity and biopsy reduction rates were calculated by applying the A6702 optimal (ADC, 1.53 × 10(-3) mm(2)/sec) and alternate conservative (1.68 × 10(-3) mm(2)/sec) cutoffs. Lesion pathologic outcomes were the reference standard. To assess reproducibility, one radiologist repeated ADC measurements, and agreement was summarized using the intraclass correlation coefficient. Results A total of 240 lesions in 201 women (mean age, 49 years ± 13 [SD]) with pathologic outcomes (63 malignant and 177 benign) were included. Applying the optimal ADC cutoff produced an overall biopsy reduction rate of 15.8% (38 of 240 lesions [95% CI: 11.2, 20.9]), with a sensitivity of 92.1% (58 of 63 lesions [95% CI: 82.4, 97.4]; sensitivity was 97.2% [35 of 36 lesions] [95% CI: 82.7, 99.6] for invasive cancers). Results were similar for screening versus diagnostic examinations (P = .92 and .40, respectively). Sensitivity was higher for masses than for nonmass enhancements (NMEs) (100% vs 85.3%; P = .009). Applying the conservative ADC cutoff achieved a sensitivity of 95.2% (60 of 63 lesions [95% CI: 86.7, 99.0]), with a biopsy reduction rate of 10.4% (25 of 240 lesions [95% CI: 6.7, 14.5]). Repeated single-reader measurements showed good agreement with clinical ADCs (intraclass correlation coefficient, 0.72 [95% CI: 0.58, 0.81]). Conclusion This study validated the clinical use of ADC cutoffs to reduce MRI-prompted biopsies by up to 16%, with a suggested tradeoff of lowered sensitivity for in situ and microinvasive disease manifesting as NME. Clinical trial registration no. NCT02022579 © RSNA, 2024 Supplemental material is available for this article. See also the editorial by Honda and Iima in this issue.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。