An incomplete pacemaker lead fracture causing inappropriate pacing inhibition due to oversensing of the minute ventilation sensor pulses

起搏器导线不完全断裂导致过度感知分钟通气传感器脉冲,从而引起不适当的起搏抑制。

阅读:1

Abstract

Cardiac pacemakers using rate response technologies represent the effectiveness of increasing the heart rate and cardiac output during exercise. Minute ventilation (MV) sensors are popular and estimate rates using transthoracic impedance by emitting very low amperage short electrical current pulses between the pacemaker lead tip and pulse generator. We present a case of an incomplete pacemaker lead fracture developing inappropriate pacing inhibition due to oversensing caused by the electrical current emitted by the MV sensor. A permanent pacemaker replacement was performed, resulting in no further abnormal findings such as inappropriate pacing inhibition. .