Comparative analysis of salivary cortisol measurements using different assay methods in relation to serum-free cortisol measurement

采用不同检测方法测定唾液皮质醇与血清游离皮质醇的比较分析

阅读:1

Abstract

OBJECTIVES: Salivary cortisol reflects the biologically active form of serum cortisol, offering a noninvasive evaluation method for the diurnal rhythm of the hypothalamic-pituitary-adrenal (HPA) axis. While liquid chromatography-tandem mass spectrometry (LC-MS/MS) is known for its specificity, immunoassays (IA) are commonly used because of their simplicity. This study aimed to assess the performance of salivary cortisol measurement using both IA and LC-MS/MS in comparison to serum-free cortisol measurement. METHODS: Assay results for 188 saliva and 94 serum samples from 47 participants were analyzed. Salivary samples collected at different time points were analyzed using IA and LC-MS/MS. Serum samples were analyzed for cortisol, cortisol-binding globulin, and free cortisol. The statistical analyses included correlations and method comparisons. RESULTS: The diurnal salivary cortisol profiles exhibited a comparable circadian rhythm pattern; however, the concentrations measured using IA were consistently higher than those measured using LC-MS/MS. The correlation analysis revealed robust associations among salivary cortisol (IA), salivary cortisol (LC-MS/MS), and serum-free cortisol levels (LC-MS/MS). However, the method comparison revealed a systematic bias between IA and LC-MS/MS in salivary cortisol measurement. CONCLUSIONS: This study contributes to the ongoing debate on assay techniques by affirming the suitability of IA and LC-MS/MS for salivary cortisol measurement to assess dynamic changes in HPA axis activity. The identified systematic bias emphasizes the importance of selecting methods based on specific research or clinical requirements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。