Enhancing optoacoustic mesoscopy through calibration-based iterative reconstruction

通过基于校准的迭代重建增强光声介观

阅读:7
作者:Urs A T Hofmann, Weiye Li, Xosé Luís Deán-Ben, Pavel Subochev, Héctor Estrada, Daniel Razansky

Abstract

Optoacoustic mesoscopy combines rich optical absorption contrast with high spatial resolution at tissue depths beyond reach for microscopic techniques employing focused light excitation. The mesoscopic imaging performance is commonly hindered by the use of inaccurate delay-and-sum reconstruction approaches and idealized modeling assumptions. In principle, image reconstruction performance could be enhanced by simulating the optoacoustic signal generation, propagation, and detection path. However, for most realistic experimental scenarios, the underlying total impulse response (TIR) cannot be accurately modelled. Here we propose to capture the TIR by scanning of a sub-resolution sized absorber. Significant improvement of spatial resolution and depth uniformity is demonstrated over 3 mm range, outperforming delay-and-sum and model-based reconstruction implementations. Reconstruction performance is validated by imaging subcutaneous murine vasculature and human skin in vivo. The proposed experimental calibration and reconstruction paradigm facilitates quantitative inversions while averting complex physics-based simulations. It can readily be applied to other imaging modalities employing TIR-based reconstructions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。