The influence of topical anesthetic and fluorescein on non-contact tonometry measurements using ultra-high-speed dynamic Scheimpflug

局部麻醉剂和荧光素对使用超高速动态Scheimpflug技术进行非接触式眼压测量的影响

阅读:1

Abstract

This study aimed to investigate the effects of topical anesthetic and fluorescein drops on intraocular pressure (IOP), central corneal thickness (CCT) and biomechanical properties as measured by Corvis ST (CST-Oculus; Wezlar, Germany) in healthy eyes. A cross-sectional observational study was conducted on 46 healthy patients. The CST measurements were obtained before and immediately after the instillation of topical anesthetic and fluorescein drops. Pre-post instillation data were statistically analyzed. IOP measurements were compared to Goldmann's Applanation Tonometry (GAT), which was also performed after drops instillation. Biomechanical parameters analyzed included applanation 1 velocity, applanation 2 velocity, applanation 1 time, applanation 2 time, whole eye movement, deflection amplitude, and stiffness parameter at first applanation. A statistically significant difference in IOP, both for non-corrected IOP (IOPnct) and biomechanically corrected IOP (bIOP), was observed before and after the instillation of eyedrops. Despite this statistical significance, the observed difference lacked clinical relevance. The IOPnct demonstrated a significant difference pre and post-anesthetic and fluorescein instillation compared to GAT (14.99 ± 2.27 mmHg pre-instillation and 14.62 ± 2.50 mmHg post-instillation, versus 13.98 ± 2.04 mmHg, with p-values of 0.0014 and 0.0490, respectively). Comparable findings were noted when justaposing bIOP to GAT (14.53 ± 2.10 mmHg pre-instillation and 13.15 ± 2.25 mmHg post-instillation, against 13.98 ± 2.04 mmHg, with p-values of 0.0391 and 0.0022, respectively). Additionally, CCT measurements revealed a statistically significant elevation following the administration of topical anesthetic and fluorescein drops (from 544.64 ± 39.85 µm to 586.74 ± 41.71 µm, p < 0.01. None of the analyzed biomechanical parameters showed statistically significant differences after drops instillation. While the administration of topical anesthetic and fluorescein drops induced a statistically significant alteration in both IOPnct and bIOP readings, these changes were not clinically consequential. Furthermore, a notable statistical rise was observed in CCT measurements post-drops instillation, as determined by CST. Yet, corneal biomechanical parameters remained unaffected.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。