Force level of small diameter nickel-titanium orthodontic wires ligated with different methods

不同结扎方法对小直径镍钛丝受力的影响

阅读:7
作者:Rodrigo Hitoshi Higa, José Fernando Castanha Henriques, Guilherme Janson, Murilo Matias, Karina Maria Salvatore de Freitas, Fernanda Pinelli Henriques, Manoela Fávaro Francisconi

Background

The

Conclusions

The SLB and the ML exhibit standard force patterns during unloading, while the elastomeric ligatures exhibit a randomly distributed force release behavior.

Methods

elastomeric ligature (EL) and metal ligature (ML). Methods: Two wire diameters (0.014 and 0.016 in.) and 10 specimens per group were used. The specimens were assembled in a clinical simulation device and tested in an Instron Universal Testing Machine, with a load cell of 10 N. For the testing procedures, the acrylic block representative of the right maxillary central incisor was palatally moved, with readings of the force at 0.5, 1, 2, and 3 mm, at a constant speed of 2 mm/min and temperature of 36.5 °C.

Results

The conventional NiTi released higher forces than the thermally activated NiTi archwires in large deflections. In general, the SLB showed lower forces, while the ML had higher forces, with both showing a similar force release behavior, constantly decreasing as the deflection decreased. The EL showed an irregular behavior. The active SLB showed smaller forces than passive, in large deflections. Conclusions: The SLB and the ML exhibit standard force patterns during unloading, while the elastomeric ligatures exhibit a randomly distributed force release behavior.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。