Encoding scheme design for gradient-free, nonlinear projection imaging using Bloch-Siegert RF spatial encoding in a low-field, open MRI system

在低场开放式磁共振成像系统中,采用Bloch-Siegert射频空间编码设计用于无梯度非线性投影成像的编码方案

阅读:1

Abstract

Eliminating conventional pulsed B(0)-gradient coils for magnetic resonance imaging (MRI) can significantly reduce the cost of and increase access to these devices. Phase shifts induced by the Bloch-Siegert shift effect have been proposed as a means for gradient-free, RF spatial encoding for low-field MR imaging. However, nonlinear phasor patterns like those generated from loop coils have not been systematically studied in the context of 2D spatial encoding. This work presents an optimization algorithm to select an efficient encoding trajectory among the nonlinear patterns achievable with a given hardware setup. Performance of encoding trajectories or projections was evaluated through simulated and experimental image reconstructions. Results show that the encodings schemes designed by this algorithm provide more efficient spatial encoding than comparison encoding sets, and the method produces images with the predicted spatial resolution and minimal artifacts. Overall, the work demonstrates the feasibility of performing 2D gradient-free, low-field imaging using the Bloch-Siegert shift which is an important step towards creating low-cost, point-of-care MR systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。