Hydrodynamic function and spring constant calibration of FluidFM micropipette cantilevers

FluidFM微移液管悬臂梁的流体动力学功能和弹簧常数校准

阅读:1

Abstract

Fluidic force microscopy (FluidFM) fuses the force sensitivity of atomic force microscopy with the manipulation capabilities of microfluidics by using microfabricated cantilevers with embedded fluidic channels. This innovation initiated new research and development directions in biology, biophysics, and material science. To acquire reliable and reproducible data, the calibration of the force sensor is crucial. Importantly, the hollow FluidFM cantilevers contain a row of parallel pillars inside a rectangular beam. The precise spring constant calibration of the internally structured cantilever is far from trivial, and existing methods generally assume simplifications that are not applicable to these special types of cantilevers. In addition, the Sader method, which is currently implemented by the FluidFM community, relies on the precise measurement of the quality factor, which renders the calibration of the spring constant sensitive to noise. In this study, the hydrodynamic function of these special types of hollow cantilevers was experimentally determined with different instruments. Based on the hydrodynamic function, a novel spring constant calibration method was adapted, which relied only on the two resonance frequencies of the cantilever, measured in air and in a liquid. Based on these results, our proposed method can be successfully used for the reliable, noise-free calibration of hollow FluidFM cantilevers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。