Index-level fusion and adjacent segment disease following dynamic stabilization for lumbar degenerative disc disease: illustrative case

腰椎退行性椎间盘疾病动态稳定术后出现节段融合及邻近节段病变:病例分析

阅读:1

Abstract

BACKGROUND: Dynesys dynamic stabilization (DDS) is an alternative to surgical fusion for the operative management of degenerative lumbar spondylosis. Compared to rigid instrumentation and fusion, DDS is purported to preserve a higher degree of spinal range of motion and reduce the risk of developing adjacent segment disease (ASD). OBSERVATIONS: A 60-year-old female presented with severe back pain and bilateral leg pain, which had progressed over the prior 4 years. Nine years earlier, she had undergone DDS system implantation at L5-S1 for lumbar stenosis and spondylosis. Repeat imaging revealed an unintended fusion at the index level (L5-S1) and ASD causing severe lateral recess stenosis at L4-5. She underwent DDS system removal, decompression at L4-5, and extension of the fusion to L4. LESSONS: Although DDS has been marketed as a motion-preserving system that avoids fusion and reduces the risk of ASD, unintended index-level fusion and ASD can still occur after DDS system surgery. These potential complications should be assessed when determining the optimal primary surgical treatment for patients with lumbar degenerative disc disease. https://thejns.org/doi/10.3171/CASE24179.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。