Impact of measurement location on direct mitral regurgitation quantification using four-dimensional flow cardiovascular magnetic resonance

测量位置对使用四维流心血管磁共振直接定量二尖瓣反流的影响

阅读:1

Abstract

BACKGROUND: Four-dimensional (4D) flow cardiovascular magnetic resonance (CMR) shows promise for quantifying mitral regurgitation (MR) by allowing for direct regurgitant volume (RVol) measurement using a plane precisely placed at the MR jet. However, the ideal location of a measurement plane remains unclear. This study aims to systematically examine how varying measurement locations affect RVol quantification and determine the optimal location using the momentum conservation principle of a free jet. METHODS: Patients diagnosed with MR by transthoracic echocardiography (TTE) and scheduled for CMR were prospectively recruited. Regurgitant jet flow volume (RVol(jet)) and regurgitant jet flow momentum (RMom(jet)) were quantified using 4D flow CMR at seven locations along the jet axis, x. The reference plane (mid-plane, x = 0 mm) was positioned at the peak velocity of the jet at each cardiac phase, and three additional planes were positioned on either side of the jet, each 2.5 mm apart. RVol(jet) was compared to RVol(TTE), measured by the proximal isovelocity surface area method, and RVol(indirect), measured by subtracting aortic forward flow volume from the left ventricle stroke volume derived from two-dimensional phase contrast at the aortic valve and a stack of short-axis cine CMR techniques. RESULTS: RVol(jet) and RMom(jet) were quantified in 45 patients (age 63±13, male 26). In patients with RVol(jet) at x = 0 mm ≥ 10 mL (n = 25), RVol(jet) consistently increased as the plane moved downstream. RVol(jet) measured furthest upstream (x = -7.5 mm) was significantly lower (39±11%, p<0.001) and RVol(jet) measured furthest downstream (x = 7.5 mm) was significantly higher (16±19%, p<0.001) than RVol(jet) at x = 0 mm. RMom(jet) similarly increased from x = -7.5 to 0 mm (57±12%, p<0.001) but stabilized from x = 0-7.5 mm (-2±17%). From x = -7.5 to 7.5 mm, RVol(jet) was in consistent moderate agreement with RVol(indirect) (n = 41, bias = -2±24 to 8±32 mL, intraclass correlation coefficient = 0.55-0.63, p<0.001). CONCLUSION: The location of a measurement plane significantly influences RVol quantification using the direct 4D flow CMR approach. Based on the converging profile of RMom(jet), we propose the peak velocity of the jet as the optimal position.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。