A novel moving phantom insert for image quality assessment in magnetic resonance imaging

一种用于磁共振成像图像质量评估的新型移动式体模插件

阅读:1

Abstract

Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.