The Degradation of Abdominal Aortic Prosthesis After 37 Years

腹主动脉人工血管使用37年后的降解

阅读:1

Abstract

An 85-year-old man underwent abdominal aortic aneurysm (AAA) repair with a synthetic graft at age 48. Sac enlargement, likely due to an endoleak, was observed during follow-up, and covered stents were placed in the graft limbs at age 80. Despite this, the sac continued to enlarge, leading to a referral to our institution. Imaging revealed an endoleak from an uncovered graft section. An EXCLUDER® stent graft (W. L. Gore & Associates, Inc., Flagstaff, AZ) was placed to cover the graft entirely, resolving the endoleak. The patient was discharged on postoperative day 5. This case highlights the rare failure of a synthetic graft after 37 years.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。