Rapid and Simple Buffer Exchange Using Cation-Exchange Chromatography to Improve Point-of-Care Detection of Pharmacological Agents

使用阳离子交换色谱法进行快速简便的缓冲液交换,以改善药物的即时检测

阅读:2
作者:Michael C Brothers, Maegan Kornexl, Barlow Guess, Yuri Kim, Darrin Ott, Jennifer A Martin, Dara Regn, Steve S Kim

Abstract

The current COVID-19 pandemic has highlighted the power, speed, and simplicity of point-of-care (POC) diagnostics. POC diagnostics are available for a wide range of targets, including both drugs of abuse as well as performance-enhancing drugs. For pharmacological monitoring, minimally invasive fluids such as urine and saliva are commonly sampled. However, false positives or negatives caused by interfering agents excreted in these matrices may confound results. For example, false positives have, in most cases, prevented the use of POC diagnostics for pharmacological agent detection; the consequence is that centralized labs are instead tasked to perform these screenings, resulting in significant delays between sampling and testing. Thus, a rapid, simple, and inexpensive methodology for sample purification is required for the POC to reach a field-deployable tool for the pharmacological human health and performance assessments. Buffer exchange is a simple, rapid approach to remove interfering agents, but has traditionally been difficult to perform on small pharmacological molecules. Therefore, in this communication, we use salbutamol, a performance-enhancing drug, as a case example to demonstrate the efficacy of ion-exchange chromatography as a technique to perform buffer exchange for charged pharmacological agents. This manuscript demonstrates the efficacy of this technique leveraging a commercial spin column to remove interfering agents found in simulant urines, such as proteins, creatinine, and urea, while retaining salbutamol. The utility and efficacy of the method was then confirmed in actual saliva samples. The eluent was then collected and run on the lateral flow assays (LFAs), improving the reported limit of detection by over 5× (new lower limit of detection of 10 ppb compared to reported 60 ppb by the manufacturer) while simultaneously removing noise due to background interfering agents.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。