Lead Break during Extraction: Predisposing Factors and Impact on Procedure Complexity and Outcome: Analysis of 3825 Procedures

拔牙过程中导线断裂:诱发因素及其对手术复杂性和结果的影响:3825例手术分析

阅读:1

Abstract

Background: Currently, there are no reports describing lead break (LB) during transvenous lead extraction (TLE). Methods: This study conducted a retrospective analysis of 3825 consecutive TLEs using mechanical sheaths. Results: Fracture of the lead, defined as LB, with a long lead fragment (LF) occurred in 2.48%, LB with a short LF in 1.20%, LB with the tip of the lead in 1.78%, and LB with loss of a free-floating LF in 0.57% of cases. In total, extractions with LB occurred in 6.04% of the cases studied. In cases in which the lead remnant comprises more than the tip only, there was a 50.31% chance of removing the lead fragment in its entirety and an 18.41% chance of significantly reducing its length (to less than 4 cm). Risk factors for LB are similar to those for major complications and increased procedure complexity, including long lead dwell time [OR = 1.018], a higher LV ejection fraction, multiple previous CIED-related procedures, and the extraction of passive fixation leads. The LECOM and LED scores also exhibit a high predictive value. All forms of LB were associated with increased procedure complexity and major complications (9.96 vs. 1.53%). There was no incidence of procedure-related death among such patients, and LB did not affect the survival statistics after TLE. Conclusions: LB during TLE occurs in 6.04% of procedures, and this predictable difficulty increases procedure complexity and the risk of major complications. Thus, the possibility of LB should be taken into account when planning the lead extraction strategy and its associated training.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。