Liquid resistivity of pharmaceutical propellants using novel resistivity cell

利用新型电阻率电池测定药物推进剂的液体电阻率

阅读:1

Abstract

Metered-dose inhalers employ propellants to produce pharmaceutical aerosols for treating respiratory conditions like asthma. In the liquid phase, the DC volume resistivity of pharmaceutical propellants, including R134a, R152a, and R227ea, was studied at saturation pressures and room temperature (not vapour phase). These measurements are essential for industries like refrigerants. Aerosols from metered dose inhalers (MDIs) with these propellants become electrically charged, affecting medicament deposition in lung. The resistivity was measured using a novel concentric cylinder-type capacitance cell designed in-house. The resistivity for the propellants (R134a, R152a, and R227ea) was found to be 3.02 × 10(10) Ωm, 2.37 × 10(9) Ωm and 1.31 × 10(10) Ωm, respectively. The electrical resistivity data obtained was found to be at least two orders of magnitude higher than the limited data available in the literature. Challenges in the resistivity cell's development and performance are discussed, with a focus on various propellants and their mixtures with ethanol and moisture concentrations. The resistivity of propellant mixtures containing moisture concentrations ranging from 5 to 500 ppm and ethanol concentrations ranging between 1000 and 125,000 ppm was determined. The resistivity was tested across 10-min and 1-h periods and was performed in accordance with the contemporary IEC 60247 standard.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。