Conclusion
These results suggest that the posterior approach LinQ Implant System is a safe and effective treatment for sacroiliac joint dysfunction at 12 months, with results that are favorable compared to outcomes reported for an FDA-cleared lateral approach.
Methods
A multi-center, prospective, single-arm study was launched after initial studies provided preliminary data to support safety, efficacy, and durability of this minimally invasive sacroiliac posterior fusion LinQ allograft implant (NCT04423120). Preliminary
Results
One-hundred and fifty-nine (159) participants were enrolled across 16 investigational sites in the US between January 2020 and March 2022. One-hundred and twenty-two (122) participants were implanted. At the 1-month follow-up, 82 participants satisfied all criteria for the composite responder endpoint, representing 73.2% of the study cohort. These results stayed consistent across the remaining study timepoints with 66.0%, 74.4%, and 73.5% of participants classified as responders at the 3-, 6- and 12-month follow-up visits, respectively. VAS scores were significantly reduced (p < 0.0001) and ODI scores were significantly improved (p < 0.0001). All domains of the PROMIS-29 were also significantly improved (all p's <0.0001). Only one procedure-related serious AE was reported in the study.