Disposable Polyaniline/m-Phenylenediamine-Based Electrochemical Lactate Biosensor for Early Sepsis Diagnosis

用于早期脓毒症诊断的一次性聚苯胺/间苯二胺基电化学乳酸生物传感器

阅读:1

Abstract

Lactate serves as a crucial biomarker that indicates sepsis assessment in critically ill patients. A rapid, accurate, and portable analytical device for lactate detection is required. This work developed a stepwise polyurethane-polyaniline-m-phenylenediamine via a layer-by-layer based electrochemical biosensor, using a screen-printed gold electrode for lactate determination in blood samples. The developed lactate biosensor was electrochemically fabricated with layers of m-phenylenediamine, polyaniline, a crosslinking of a small amount of lactate oxidase via glutaraldehyde, and polyurethane as an outer membrane. The lactate determination using amperometry revealed the biosensor's performance with a wide linear range of 0.20-5.0 mmol L(-1), a sensitivity of 12.17 ± 0.02 µA·mmol(-1)·L·cm(-2), and a detection limit of 7.9 µmol L(-1). The developed biosensor exhibited a fast response time of 5 s, high selectivity, excellent long-term storage stability over 10 weeks, and good reproducibility with 3.74% RSD. Additionally, the determination of lactate in human blood plasma using the developed lactate biosensor was examined. The results were in agreement with the enzymatic colorimetric gold standard method (p > 0.05). Our developed biosensor provides efficiency, reliability, and is a great potential tool for advancing lactate point-of-care testing applications in the early diagnosis of sepsis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。