Evaluating Factors Affecting Mean Glandular Dose in Mammography: Insights from a Retrospective Study in Dubai

评估影响乳腺X线摄影平均腺体剂量的因素:来自迪拜一项回顾性研究的启示

阅读:1

Abstract

BACKGROUND/OBJECTIVE: This study evaluates the mean glandular dose (MGD) in mammography screening for women aged 40-69 in Dubai, based on a retrospective analysis of a dose survey involving 2599 participants. METHODS: MGD was calculated using the Dance formula. RESULTS: The average MGD was 0.96 ± 0.39 mGy for mediolateral oblique (MLO) views and 0.81 ± 0.33 mGy for craniocaudal (CC) views. Weak inverse correlations were found between age and organ dose (OD) for both views, while a direct relationship was observed between breast thickness and entrance skin dose (ESD). In adjusted models, ESD was strongly associated with MGD (β = 1.04, 95% CI: 0.97, 1.09), while OD showed a moderate association (β = 0.44, 95% CI: 0.40, 0.49). Significant variations in ESD, OD, and MGD were noted across age groups and breast thicknesses. CONCLUSIONS: Lower MGD indicates reduced radiation exposure risk, while higher MGD in MLO views suggests improved imaging quality. Monitoring and optimizing MGD are essential for enhancing patient safety and screening efficacy.